K. Tsen: UV Treatment of Blood vs Cancer -- US Patent #
6113566

![](0logo.gif)  
**[rexresearch.com](../index.htm)**

---

**K.T. TSEN, *et al*:**

**Inactivation of Viruses by UV Light**

---

**<http://www.newscientisttech.com/article/dn12368-visible-light-pulses-knock-out-viruses-in-blood.html>**

**"Visible Light Pulses Knock Out Viruses
in Blood"**

**by** **Belle Dume**

The technique destroys a virus with a pulse of light from a
low-power laser. The pulse produces mechanical vibrations in the
virus shell, or capsid, irreversibly damaging and disintegrating
it, and so "deactivating" the virus for good. The technique
might be used to kill HIV, as well as hepatitis C, say the
researchers involved. Traditional methods of destroying viruses,
such as UV irradiation, can cause mutations, which eventually
make the micro-organisms resistant. UV light can also damage the
DNA of surrounding healthy cells. Scientists have also tried
using microwaves to kill viruses but this is even less promising
since the water in and around a micro-organism strongly absorbs
this frequency of light. Most of the energy from the microwave
radiation is absorbed by the water and does not even reach the
virus itself. The researchers applied pulses of purple-coloured
light lasting just 100 femtoseconds (10-15 seconds) to viruses
called M13 bacteriophages. It takes just a single pulse to
destroy the viruses completely, say the researchers. The "power
density" of the laser is just 50 megawatts per square
centimetre, which is low enough to leave surrounding human cells
and tissue undamaged, but high enough to produce large-amplitude
vibrations in a virus's capsid. It is also too low to cause
genetic mutations, meaning the virus will not build up resistant
to the treatment over time. Disinfecting blood - Tsen told New
Scientist that the technique could be used to disinfect blood or
other biological samples in hospitals. "In addition, we believe
that the method may be especially important in designing novel
treatments for blood-borne viral diseases," he said. "For
example blood dialysis allows us to irradiate a patient's blood
outside the body and potentially cleanse it of infectious virus
particles before reintroducing it into the patient. In this way,
we could reduce mortality associated with diseases like
hepatitis C and AIDS." The team now plans to test the efficacy
of its technique in killing a wide range of deadly viruses,
including HIV and hepatitis C. "We also plan to conduct further
tests on the effects of the low-power visible laser on mammalian
cells to determine any potential side effects and confirm that
it selectively kills viruses," said Tsen.

---

**http://www.iop.org/EJ/abstract/0953-8984/19/32/322102**

**Inactivation of Viruses with a Very Low Power Visible
Femtosecond Laser**

**K T Tsen, et al., 2007 *J. Phys.: Condens. Matter* 19,**
322102 (9pp); doi: 10.1088/0953-8984/19/32/322102

K T Tsen1, Shaw-Wei D Tsen (2), Chih-Long Chang (2), Chien-Fu
Hung (2,3), T-C Wu (2,3,4,5) and Juliann G Kiang (6,7,8)

(1)  Department of Physics, Arizona State University,
Tempe, AZ 85287, USA   
(2)  Department of Pathology, Johns Hopkins School of
Medicine, Baltimore, MD 21231, USA   
(3)  Department of Oncology, Johns Hopkins School of
Medicine, Baltimore, MD 21231, USA   
(4)  Department of Obstetrics and Gynecology, Johns Hopkins
School of Medicine, Baltimore, MD 21231, USA   
(5)  Department of Molecular Microbiology and Immunology,
Johns Hopkins School of Medicine, Baltimore, MD 21231, USA   
(6)  Scientific Research Department, Armed Forces
Radiobiology Research Institute, Uniformed Services University
of The Health Sciences, Bethesda, MD 20889-5603, USA   
(7)  Department of Medicine, Uniformed Services University
of The Health Sciences, Bethesda, MD 20889-5603, USA   
(8)  Department of Pharmacology, Uniformed Services
University of The Health Sciences, Bethesda, MD 20889-5603, USA

**Abstract --** We demonstrate for the first time that, by
using a visible femtosecond laser, it is effective to inactivate
viruses such as bacteriophage M13 through impulsive stimulated
Raman scattering. By using a very low power visible femtosecond
laser having a wavelength of 425 nm and a pulse width of 100 fs,
we show that M13 phages were inactivated when the laser power
density was greater than or equal to 50 MW cm-2. The
inactivation of M13 phages was determined by plaque counts and
depended on the pulse width as well as power density of the
excitation laser.

---



**US Patent # 6,113,566**

**Ultraviolet Blood Irradiation Method and
Apparatus**

**Carl Schleicher**

**Abstract ---** Ultraviolet light produces a rapid
detoxifying effect with subsidence of toxic symptoms. Venous
oxygen is increased in patients with depressed blood oxygen
values. Of special interest is that a rapid rise in resistance
to acute or chronic viral and bacterial infection occurs. No
harmful affects have been observed with UBI therapy in thousands
of cases of viral infections, hepatitis, bacterial infections,
hypoxemia and many other illnesses, especially blood-related
infections. The diseases successfully treated with UBI include:
(1) Atypical pneumonia; (2) Poliomyelitis and polioencephalitis;
(3) Hepatitis; infectious and serum; (4) Influenza; (5) common
upper respiratory disease; (6) Herpes simplex; (7) Herpes
zoster; (8) Mumps; (9) Mononucleosis; and (10) measles.
Moreover, preliminary reports indicate that UBI may be useful in
treating HIV and research is currently under way to evaluate the
effects of UBI on eliminating HIV from blood and blood products.
If this research is successful, it would have major implications
in ensuring the safety of blood in blood banks.

Schleicher; Carl (Silver Spring, MD)   
Assignee:  Foundation for Blood Irradiation Inc. (Silver
Spring, MD)

Current U.S. Class:  604/6.08 ; 607/92   
Current International Class:  A61M 1/36 (20060101); A61M
037/00 ()   
Field of Search:  604/4 607/88-94
250/432R,455.11,435,436,437,438 422/24,44   
References Cited : U.S. Patent Documents -- 1200940 
October 1916  Henri et al.   
 1683877  September 1928  Edblom et al. // 
2074909  March 1937  Herzig et al.   
 2308516  January 1943  Knott // 
2309124  January 1943  Knott   
 3926556  December 1975  Boucher   
 4321919  March 1982  Edelson //   
 4398906  August 1983  Edelson //   
 4428744  January 1984  Edelson //   
 4464166  August 1984  Edelson //   
 4612007  September 1986  Edelson //   
 4613322  September 1986  Edelson //   
 4683889  August 1987  Edelson //   
 4684521  August 1987  Edelson //   
 4737140  April 1988  Lee et al. //   
 5150705  September 1992  Stinson //   
 5263925  November 1993  Gilmore et al. //   
 5290221  March 1994  Wolf, Jr. et al. //   
 5429594  July 1995  Castle //   
 5433738  July 1995  Stinson //

**Description**

**BACKGROUND OF THE INVENTION**

**1. Field of the Invention**

The invention relates to both method and apparatus for
irradiating blood with ultraviolet radiation, and more
particularly to such method and apparatus for in vitro radiation
of blood from both humans and animals with stored blood or blood
circulated outside the patient and returned to the patient after
treatment.

**2. Related Art**

The present status of blood irradiation treatment is
exemplified by the following patents.

(1) U.S. Pat. No. 1,200,940; Henri et al.; "Apparatus for
treatment of Water and Other Liquids by Ultra Violet Rays:
wherein water is sterilized by exposure to ultra-violet rays
from a source above or within the water.

(2) U.S. Pat. No. 1,683,877; Edblom et al.; "Means for Treating
Blood Stream Infections", wherein the apparatus used is a
veni-puncture needle inserted into the patient's vein and the
butt end of the needle is attached to tubing into which a
container of citrate or oxylate solution is connected. A
transfusion pump is attached to the other end of the tubing and
the outlet of the pump is connected to an exposure chamber
covered by a quartz lens. The source generates rays between
1800-4000 Angstroms.

(3) **U.S. Pat. No. 2,074,909; Herzig et al.; "Activation
Device for the Heliopyretic Treatment of Matter"**; wherein
a plurality of parallel radio-active penetrable tubes mounted in
a frame are adapted to convey fluid to be exposed to the rays
from quartz tubes and a reflector adjustable mounted in the
frame to focus the rays.

(4) **U.S. Pat. No. 2,308,516; Knott; "Method and Means for
Irradiating Blood"**; wherein a quantity of blood is removed
from the patient, passed through a chamber where the blood is
exposed to ultra-violet radiation for a brief, accurately
controlled time and returned to said circulatory system.

(5) **U.S. Pat. No. 2,309,124; Knott; "Ultra-Violet Exposure
Chamber"**; wherein a chambered receptacle having a series
of passageways extending transversely of the chamber and a
series of spirally twisted devices lie centrally of the
transverse passageways to produce turbulence in the blood flow
through the chambered receptacle.

(6) **U.S. Pat. No. 4,321,919; Edelson; "Method and Apparatus
for Externally Treating Human Blood"**; wherein in a method
for reducing the lymphocyte population by withdrawing blood from
the subject, flowing the blood stream through a treatment
station transparent to UV radiation and irradiating the blood
stream in the treatment station with UV radiation in the
presence of about 1 nanogram to micrograms of a dissolved
psoralen capable of bonding the psoralen and the lymphocytes.

(7) **U.S. Pat. No. 4,398,906; Edelson; "Method of Externally
Treating the Blood"**; wherein the functioning population of
nucleated cells in the blood supply of a human subject by
withdrawing blood from the subject and irradiating the withdrawn
blood with UV radiation in the presence of an effective amount
of a dissolved chemical agent having an affinity for the nucleic
acid of the nucleated blood cells to form photo-adducts with DNA
to thereby effect chemical bonding between the photoactive
chemical agent and the nucleic acid of the nucleated cells.

(8) **U.S. Pat. No. 4,464,166; Edelson; "Method for
Externally Treating the Blood"**; wherein the functioning
population of a nucleated cell in the blood supply of a human
subject by withdrawing blood from the subject and irradiating
the withdrawn blood with UV radiation in the presence of a
dissolved photoactivated antibody specific for the nucleated
blood cell to form photoadducts with the nucleated blood cells
to thereby effect chemical bonding between the photoactivated
antibody and the nucleated cells.

(9) **U.S. Pat. No. 4,612,007; Edelson; "Method and System
for Externally Treating the Blood"**; wherein the
functioning population of a blood constituent in the blood
supply of a human subject by irradiating withdrawn blood by UV
radiation in the presence of an effective amount of a dissolved
photoactive chemical agent specific for a receptor site in the
blood constituent and capable when activated by UV radiation of
forming photo-adducts with blood constituent receptor sites, to
thereby effect chemical bonding between the photoactivated
chemical agent and the receptor sites.

(10) **U.S. Pat. No. 4,613,322; Edelson; "Method and System
for Externally Treating the Blood"**; wherein the
functioning population of a blood constituent in the blood
supply of a human subject is reduced by treating withdrawn blood
by subjecting it to a density gradient to effect a substantial
separation of the red blood cells from the blood constituent
sought to be reduced in population and irradiating the treated
blood with UV radiation so that the radiation impinges upon the
constituent in the presence of an effective amount of a
dissolved photoactive chemical agent specific and capable when
activated by the UV radiation of forming photo-adducts with
blood constituent receptor sites, to thereby effect chemical
bonding between the photoactivated chemical agent and the
receptor sites.

(11) **U.S. Pat. No. 4,683,889; Edelson; "Method and System
for Externally Treating the Blood"**; wherein in a system
for treating blood from a subject to reduce the functioning
population of a blood constituent by associating and reacting
the constituent with a photoactive agent including means for
receiving blood and subjecting it to a density gradient for
separating red blood cells from the blood constituents, and
preferentially impinging UV radiation upon the blood constituent
to react an associated photoactive agent with the blood
constituent.

(12) **U.S. Pat. No. 4,684,521; Edelson; "Method and System
for Externally Treating the Blood"**; wherein a chemical
agent that is useful for reducing the population of a selected
blood constituent having receptor sites comprises a carrier
having a strong affinity for the receptor sites on or in the
selected blood constituents and a photoactive agent physically
incorporated within or chemically bound to the carrier moiety to
interfere with the metabolism of the selected blood constituent
when activated with UV radiation.

(13) **U.S. Pat. No. 4,737,140; Lee et al.; "Irradiation
Chamber for Photoactivation Patient Treatment System"**;
wherein a removable UV light ray assembly for use in a
photoactivatable agent, patient treating system wherein
photoactivatable agents in contact with patient blood cells are
irradiated extracorporeally and then returned to the patient.

(14) **U.S. Pat. No. 5,150,705; Stinson; "Apparatus and
Method for Irradiating Cells"**; wherein an apparatus for
irradiating cells with UV light includes a UV light source and
an outer cylinder surrounding the UV light source with a hollow
tubing wrapped around the outer peripheral surface of the outer
cylinder whereby suspended cells are transported within the
hollow tubing so that the cells can be irradiated by the UV
light source.

(15) **U.S. Pat. No. 5,263,925; Gilmore, Jr. et al.;
"Photopheresis Blood Treatment"**; wherein portable
apparatus is worn by the patient and blood circulating within a
treatment station is irradiated with rays with continuous blood
flow through the treatment station in a closed loop path.

(16) **U.S. Pat. No. 5,290,221; Wolf, Jr. et al.; "Systems
for Eradicating Contaminants Using Photoactive Materials in
Fluids Like Blood"**; wherein the blood is constricted into
a relatively narrow, arcuately shaped flow path into a radiation
chamber that directs radiation from one or more sources into the
blood.

(17) **U.S. Pat. No. 5,429,594; Castle; "Extra-Corporeal
Blood Access, Sensing and Radiation Methods and Apparatuses"**;
wherein blood is treated with radiation in a system in which an
apparatus includes one or more access ports in tubing through
which blood flows with treatment and/or analysis windows
disposed adjacent, within or over the ports.

(18) **U.S. Pat. No. 5,433,738; Stinson; "Method for
Irradiating Cells"**; wherein a method for irradiating cells
with UV light includes a UV light source and an outer cylinder
surrounding the UV light source with a hollow tubing wrapped
around the outer peripheral surface of the outer cylinder
whereby suspended cells are transported within the hollow tubing
so that the cells can be irradiated by the UV light source. An
inner cylinder can be positioned inside the outer cylinder
between the UV light source and the outer cylinder and
ventilating means may be employed to maintain a substantially
constant temperature.

Notwithstanding the advances made in blood irradiation as
represented by the foregoing U.S. patents, there is a need for
improvements in at least the cuvette employed in the irradiation
stations, the manner of housing the pump and irradiation station
and improving the availabe range of UV wavelength output to the
irradiation station from the UV light source.

**SUMMARY OF THE INVENTION**

Ultraviolet (UV) blood irradiation (UBI) therapy may be
administered by a device called a Knott HEMO-IRRADIATOR.RTM.,
for example as disclosed in Knott's U.S. Pat. No. 2,308,516
identified above under paragraph (4). UBI therapy raises the
resistance of the host and is therefore able to control many
disease processes. A fundamental effect of UBI is to energize or
enhance the biochemical and physiological defenses of the body
by the introduction of UV energy into the blood stream. It is
well known that UV radiation is used to purify water and treat
sewage. It has been amply demonstrated, as evidenced by the
above cited patents, that UV radiation can purify and clean the
blood of contaminations in the form of viruses and bacteria.

Such treatment is intravenously applied by irradiating blood
with a controlled amount of UV energy in the accepted
therapeutic band. This produces a rapid detoxifying effect with
subsidence of toxic symptoms. Venous oxygen is increased in
patients with depressed blood oxygen values. Of special interest
is that a rapid rise in resistance to acute or chronic viral and
bacterial infection occurs. No harmful affects have been
observed with UBI therapy in thousands of cases of viral
infections, hepatitis, bacterial infections, hypoxemia and many
other illnesses, especially blood-related infections.

The diseases successfully treated with UBI include: (1)
Atypical pneumonia; (2) Poliomyelitis and polioencephalitis; (3)
Hepatitis; infectious and serum; (4) Influenza; (5) common upper
respiratory disease; (6) Herpes simplex; (7) Herpes zoster; (8)
Mumps; (9) Mononucleosis; and (10) measles.

Moreover, preliminary reports indicate that UBI may be useful
in treating HIV and research is currently under way to evaluate
the effects of UBI on eliminating HIV from blood and blood
products. If this research is successful, it would have major
implications in ensuring the safety of blood in blood banks.

The Knott technique of blood irradiation (approved by the
American Blood Irradiation Society) has achieved the following
physiologic objectives: (1) increases the blood oxygen level;
(2) increases phagocytosis; (3) relieves toxemia; (4) decreases
edema; and (5) controls nausea and vomiting.

Treatment generally consists of withdrawing from 1.0 to 1.5 cc
of blood per pound of body weight from the patient, citrating it
and, by use of the Knott HEMO-IRRADIATOR.RTM. exposing it to
radiant energy between the wave lengths of 2,000 and 12,000
angstroms units as it passes through the irradiation unit at a
predetermined flow rate. The blood is returned to the patient
through the needle used for the initial venipuncture. Treatment
requires from 30-45 minutes. Outpatients rest fifteen minutes,
after which time they may resume whatever activity is permitted.

The primary object of the present invention is to provide an
improved blood irradiation method and apparatus based on the
aforementioned Knott-HEMO-IRRADIATOR.RTM. device which has FDA
approval.

Other objects, advantages and features of the present invention
are to: (1) provide a pump and UV light source interface that
enables the UV light source to be changed without disassembling
the pump; (2) provide solid state electronic circuitry thereby
eliminating the use of vacuum tubes; (3) provide a counter to
record the number of exposure sessions, thereby making available
a usage history and assist in billing patients; (4) incorporate
power circuitry to allow both international and domestic use;
(5) use a UV radiation source which emits relatively little heat
during usage, thus eliminating the need for a cooling system and
also allowing the use of polymeric materials; (6) incorporate a
self-diagnostic capability to monitor the pump, UV source and
control system; (7) provide the ability to select the type of UV
range, i.e. UVA, UVB or UVC; (6) incorporate a data output port
for printer, floppy drive, hard drive, etc.; (9) significantly
reduce the size and weight of the apparatus for easier
portability of the apparatus; (10) use a clear polymeric
irradiation chamber, preconnected to the disposable IV set; (11)
provide use of a material for use as a cuvette for irradiating
blood and which is equivalent to quartz crystal in the transfer
of UV to the irradiated blood for inactivation of pathogens in
the blood; (12) to provide disposable cuvettes; (13) provide
total containment of the UV source and irradiation chamber to
prevent leakage of UV energy into the environment and include a
cover shield for shielding the eyes of the operator while
enabling observation of the blood flow through the cuvette in
the irradiation station; (14) a UV blood irradiation apparatus
that fully complies with the safety regulations of OSHA and
other government regulatory agencies and to incorporate sensors
to monitor UV intensity motor speed, blood flow rate, etc.; (15)
provide intermittent operation of the UV source coupled with
providing turbulent blood flow within irradiation chamber; (16)
to provide various wavelengths in the range from 2,000 to 12,000
Angstroms for the irradiation treatment of blood; (17) enable
improved quality control of the control circuitry for
controlling the pump and the UV light source; and (18) provide
apparatus of the type specified herein that is user friendly to
the average nurse or technician by providing standardized IV
sets.

**BRIEF DESCRIPTION OF THE DRAWINGS**

The above objects, features and advantages of the invention are
believed to be readily apparent from the following description
of a preferred embodiment representing the best mode for
carrying out the invention when taken in conjunction with the
following drawings, wherein:

**FIG. 1** illustrates a top perspective view of the UV
blood irradiator apparatus of the preferred embodiment of the
invention with the top cover closed as it would be during
excitation of the UV light source or non-use of the irradiating
apparatus;

![](6113-1.jpg)

**FIG. 2** illustrates the top perspective view of the
invention of FIG. 1 with the cover, pump and irradiation station
assembly in an exploded view, illustrating the inside of the
irradiation apparatus housing;

![](6113-2.jpg)

**FIG. 3** is a cross section view taken along lines 3--3 of
FIG. 2 and illustrating the mounting of the pump and irradiation
station;

![](6113-3.jpg)

**FIG. 4** is a block diagram of the control circuitry for
the pump and the UV light source;

![](6113-4.jpg)

**FIG. 5** is a block diagrammatic representation of the
circuitry for sensing current flow to the UV light source and
the pump;

![](6113-5.jpg)

**FIG. 6** shows a diagnostic monitor for receiving
diagnostic inputs from the UV light source, pump control, UV
light source, pump and control sensors and for providing control
input signals to the control circuitry of FIG. 4;

![](6113-6.jpg)

**FIG. 7** shows the digital output ports for providing
digitized UV light source, pump and blood flow signals as
generated by A/D conversion of the respective outputs of the UV
light source, pump and blood flow sensors;

![](6113-7.jpg)

**FIGS. 8A and 8B** respectively illustrate front and rear
perspective views of a first embodiment of an irradiation
cuvette for irradiating blood flowing therethrough when the
cuvette is inserted into the irradiation chamber and which is
designed to create turbulence in the blood flow; and

![](6113-8ab.jpg)

**FIGS. 9A and 9B** illustrate a second embodiment of an
irradiation cuvette for irradiating blood flowing therethrough
when the cuvette is inserted into the irradiation chamber and
which is thin to prevent blockage of the UV irradiation by the
blood flow.

![](6113-9ab.jpg)

**DETAILED DESCRIPTION**

With reference to the perspective view of the irradiation
apparatus of the invention as shown in FIG. 1, the blood
irradiation apparatus 10 includes a front panel 12 with a
display 13 and power control switches 14 comprising ON/OFF main
power switch 15, ON/OFF pump control switch 16 and ON/OFF UV
light control switch 17, thereby providing separate power
control to the irradiation device 10 itself as well as separate
power control of the pump and UV light source (both components
not shown in FIG. 1).

The housing 11 of the blood irradiation apparatus 10 further
comprises a tubular housing 18 with a partially rotatable cover
19 opening to the rear of the housing 10 as shown in FIG. 2.
Section 20 of housing 18 is secured in the body of housing 11 as
illustrated in FIG. 1. Lip 21 of section 20 and lip 22 of
rotatable cover 19 are closed on one another with cover 19 in a
closed position as illustrated in FIG. 1. Rotatable cover 19 may
be made of metal or, in a preferred embodiment of the invention,
of UV plastic. UV plastic is lighter than metal, and enables the
operator of the blood irradiator apparatus of the invention to
observe the flow of blood through a cuvette (to be described
with respect to FIG. 2) in an irradiation station 25 (FIG. 2)
located beneath cover 19. Aperture 23a in section 20 provides
egress for the conduit transporting the blood from the patient
to the pump 24 (FIG. 2).

Housing 10 and the various structural components described
above are preferably made from metal similar to that used for
storage cabinets.

The exploded perspective view of the blood irradiator apparatus
of FIG. 2 illustrates the relationship between the blood pump 24
and the irradiation station 25. A cover plate 26 (shown in FIG.
3) is removed in FIG. 2 to more clearly illustrate the pump and
the irradiation station. Brackets 35a and 35b support respective
opposite ends of the cover plate 26 as shown in FIG. 3.

Blood from the patient is transported via a conduit (see FIG.
3) through aperture 23a to the pump 24 and thence to a cuvette
(not shown) mounted in irradiation station 25 by brackets 26,
28, thereby enabling blood flowing through the cuvette to be
irradiated by the UV light from UV light source 29. Blood from
the cuvette is then transported via a conduit (see FIG. 3)
through aperture 23b and returned to the patient by means well
known to those skilled in the blood irradiation art.

With cover 19 in the open position as illustrated in FIG. 2,
access is provided to the pump 24 and the irradiation station
for purposes of removing or servicing these components, for
example to change the UV light source 29 or replace and position
the cuvette in brackets 26, 28 of the irradiation station 25.

The lower section 30 of housing 11 provides space for mounting
the electrical components of the blood irradiation apparatus 10.
Such components are illustrated by block components 31, 32 and
33. The manner in which such electrical components are stored in
lower section 30 forms no part of the present invention as one
of ordinary skill in the blood irradiation art would readily
perceive the location of such components in order to carry out
the invention.

The cross section view of FIG. 3 shows the manner in which the
pump 24 and the irradiation station 25 including UV light source
29 are mounted in lower section 20 of tubular housing 18.
Shoulder 34 of pump 24 rests on cover plate 26 and is secured
thereto by fastening means such as machine screws (not shown)
and cover plate 26 in turn is secured to bracket 35a attached to
the side of lower section 20. Cover plate 26 is similarly
secured to bracket 35b at the other end of lower section 20.
Cover plate 26 rests on support 37 which includes protrusion 38a
for supporting support member 39a upon which is secured one end
of bracket 26 of the irradiation station 25. Similarly the other
end of cover plate 26 is supported on support member 39b which
is secured to protrusion 38b. The other end of bracket 26 of the
irradiation station is secured to support member 39b by
fastening means 40 as shown in FIG. 3.

UV light source 29 is removably retained at each end by
respective holding elements 41a, 41b, each of which comprises a
support 42a and 42b having respective sockets 43a and 43b.
Electrical power is supplied to pump 24 via electrical
conductors 44, 45. The electrical conductors for supplying power
to the UV light source 29 are not shown in FIG. 3.

Tube 46a conveys blood from the human or animal patient to the
pump 24; tube 46b conveys blood from pump 24 to the cuvette in
irradiation station 25; and tube 46c returns blood from the
irradiation station 25 to the human or animal patient (not
shown).

The control circuitry 49 for controlling the UV light source 29
and the pump 24 is shown in block diagrammatic format in FIG. 4,
wherein both the UV light source 29 and pump 24 receive power
through respective ON/OFF power switches 16 and 17 from power
source 40. In a preferred embodiment of the invention, UV light
source 29 is one of several UV light sources, each one providing
a UV light output having an appropriate bandwidth to cover the
respective UVA, UVB and UVC wavelengths. Control input signals
41 from a diagnostic monitor (to be described with respect to
FIG. 7) are input to UV control circuitry 42 so that the
particular UV light source 29 maintains the proper light output
amplitude (wattage).

In a further modification of the present invention, a broad
band UV light source such as a quartz lamp known to the art
serves as the UV light source 29, and various filters each
having a different light transmission bandwidth, for example for
each of the UVA, UVB and UVC light bandwidths) are inserted
within bracket 26 of irradiation station 25 (FIG. 2) to be
positioned between the UV light source 29 and the cuvette
through which the blood is pumped in the irradiation station. In
this modified embodiment of the invention a select light source
signal 43 from a control panel 44, having, for example, three
selector switches 45, 46 and 47, on front panel 12 of the
HEMO-IRRADIATOR.RTM. apparatus 11 of the invention (FIG. 1)
provides a control input to UV control circuitry 42 to properly
control the excitation to the UV light source to obtain the
desired UV light output amplitude in accordance with the UV
light source that is selected.

In general it is possible to vary the light source intensity by
changing either the voltage or current of a light source and
thereby vary its radiant output. With incandescent sources such
variation can extend over a wide range. With arc sources the
range of variation is narrower since the arc can not be allowed
to extinguish or the UV treatment will fail. The frequency
output wavelengths can be adjusted by using filters or changing
the UV light source.

In a similar manner pump control circuitry 48 is powered from
mains power source 40 and receives a select pump speed signal 49
from pump speed select switch 50 mounted on front panel 12 of
the HEMO-IRRADIATOR.RTM. apparatus 10 illustrated in FIG. 2.
Speed select switch may comprise a rotatable potentiometer
switch to provide a continuum of pump speeds. It is known that
it is desirable that the blood flow through the cuvette in the
irradiator station 25 be varied in accordance with the type of
cuvette being used and the UV light output amplitude from the UV
light source.

For various physiological and/or operational reasons it may be
desirable to increase/decrease the blood flow rate. In the
apparatus of the present invention this variation in the blood
flow rate is provided as a manual operation (see FIG. 4) As
illustrated therein, there are two inputs, namely light source
48 and pump speed 49. These inputs are reference inputs which
are varied manually to obtain the desired blood flow rate and/or
radiation level.

Pump control circuitry also receives control input signals 51
from the diagnostic monitor 60 (to be more fully described with
respect to FIG. 6).

In HEMO-IRRADIATOR.RTM. apparatus of the type disclosed herein
it is desirable to determine the usage of the pump 24 and UV
light source 29 so that, for example the pump may be removed for
necessary maintenance and the anticipated need for replacing the
UV light source 29 anticipated as it approaches its specified
life cycle. To that end FIG. 5 illustrates the use of counters
55 and 56 which are respectively sensing the current flow to UV
light source 29 and pump 24.

UV light source sensor 61 and pump sensor 62 each respectively
sense the current and/or voltage of the UV light source 29 and
the pump 24, respectively. These analog signals may be converted
by A/D converters and then respectively compared with stored
values in, for example a ROM, in diagnostic monitor 60.
Differences between the sensed current and/or voltages and the
stored values provide respective control signals 41 and 51 to
the respective UV control circuitry 42 and the pump control
circuitry 48 of FIG. 4.

Thus, diagnostic monitor 60 senses the operation of the pump
control circuit 48 and the UV control circuit 42 of FIG. 4 to
determine if the control circuitry is working. This is important
as improper operation of the control circuitry results in a
failed UV blood irradiator treatment.

FIG. 7 shows the digital output ports for providing digitized
UV light source, pump and blood flow signals as generated by A/D
conversion of the respective outputs of the UV light source,
pump and blood flow sensors, which digitized data may be
provided at suitable ports located on the back of the
HEMO-IRRADIATOR.RTM. apparatus (not shown) for transmission via
modems by telephone and/or cable to remote locations. Thus the
respective outputs of UV light source sensor 61 and pump sensor
62 are converted by A/D converters 65 and 66 and their
respective digital outputs are provided to digital output ports.
In a similar manner, the output of blood flow sensor 69 is
converted by A/D converter 70 and provided as a digitized output
at digital port 71.

FIG. 8A is a front perspective view of first embodiment of a
cuvette 80 capable of being used in the irradiation station 25
(FIG. 2) of the blood irradiation apparatus of the subject
invention and which requires a modification of the irradiation
station to accommodate the cuvette therein, which modification
will be described hereinafter. Cuvette 80 may have an elliptical
or oval shape comprising a main body portion 81 and a lip 82
extending around the periphery of body portion 81. Lip 82
provides a means for fastening cuvette 80 in irradiation station
25 (described hereinafter). Blood flows into cuvette 80 via
inlet 83 and flows from the cuvette via outlet 84 (see FIG. 8B).

The interior of cuvette 80 comprises a number of
parallely-spaced chambers 85, 86, 87, 88 and 89 through which
blood successively flows in alternate directions (as indicated
by the directional arrows in FIG. 8A) from inlet 83 and entry
port 90 to outlet port 91 and then from cuvette 80 via outlet
84. The blood flows through chambers 85, 86, 87, 88 and 89
within respective channels 90, 91, 92, 93, 94, 95, 96 and 97.
Each of the channels includes a respective diverter 98, 99, 100,
101, 102, 103, 104 and 105 located at the end of each channel in
the direction of the blood flow and each of which interrupts the
flow of blood and causes it to drop onto an adjacent channel.

The above-described construction of cuvette 80 causes the blood
flow therein to be turbulent, thereby eliminating the tendency
of the blood to coagulate and also providing an appropriate
exposure of the blood from the UV source (not shown) but which
irradiates the blood through a quartz or polymeric facing which
covers the interior of the cuvette 80.

The rear view of cuvette 80 shown in FIG. 8B illustrates back
cover 106, main body portion 81, lip 82 and the spaced
relationship of blood inlet 83 and blood outlet 84.

The second embodiment of a cuvette illustrated in FIGS. 9A and
9B is a preferred embodiment of the invention and may be
disposable to eliminate the contamination of blood from one
blood sample to the next. Cuvette 110 is an elongated hollow
tube 112 having an oval-shaped cross-section as illustrated in
FIG. 9B and which includes a necked-down portion 113 and 114 at
the blood inlet 115 and blood outlet 116, respectively, thereby
diverting the blood flow into the respective upper and lower
halves of hollow tube 112 as shown in FIG. 9A. The Hollow tube
section A shown in FIG. 9A is approximately 5.0" long, 1.0" wide
(Dimension C), and each of inlet 115 and outlet 116 (dimension
B) is approximately 1.0".

FIG. 9B is a sectional view of the cuvette 110 taken along
lines 117--117 of FIG. 9A. As illustrated in FIG. 9B, hollow
tube 112 has inner spacing (dimension D) of approximately 0.85",
an inner width of 0.837" (dimension E) and an outer width of
approximately 0.96". With the above dimensions, hollow tube 112
has a diameter of approximately 0.0615 inches (dimension
F-dimension E) at the outer wall portions and a diameter of
approximately 0.075" at the end portions 120, 121, 122 and 123
(dimension C-dimension D).

It is desirable that the HEMO-IRRADIATOR.RTM. apparatus of the
present invention be capable of wide usage in different
countries world wide and therefore it is advantageous that the
power supply of the apparatus be capable of accepting both 110
and 220 volt power supply inputs and transform that voltage
input to 110 volts AC, the voltage required to operate the
apparatus of the invention. Those of ordinary skill in the art
of power supplies will readily recognize that such power
transformers are available for use in the present invention,
such that no further description of their operation is necessary
for the present invention to be practiced.

The above description serves only to describe exemplary
embodiments of the best mode of making the HEMO-IRRADIATOR.RTM.
apparatus of the invention and to demonstrate the features and
advantages of its construction and operation. The invention is
not intended to be limited thereby, as those skilled in the
hemo-irradiator art will readily perceive modifications of the
above-described embodiments. Thus the invention is intended to
be limited only by the following claims and the equivalents to
which the claimed components thereof are entitled.

---